With the development of molecular biology, genetic engineering research and applications, centrifugal separation technology has also undergone several generations of replacement, from low-speed centrifuges, high-speed centrifuges, ultracentrifuges to ultra-speed refrigerated centrifuges, making centrifugal separation technology a new Period. The existing analytical ultracentrifuge is also equipped with an analytical optical system that can be used for quantitative analysis. This article describes the common faults, troubleshooting methods and maintenance precautions of the centrifuge during clinical use.
In the process of use, the motor can not start normally. If the main power indicator light is on, the degree of wear of the carbon brush should be checked. If the carbon brush wear exceeds one third of the total amount, the new carbon brush should be replaced in time. When the main power indicator is off, check if the indicator fuse and the indoor distribution board fuse are blown. Also check whether the power line is in good contact. If the fuse is broken, replace the fuse and keep the circuit open. Also check the vacuum pump gauge and oil pressure indication value. If the oil pressure is too high and the main engine cannot start, it is necessary to check whether the oil passages are blocked, especially if the throttle orifice is unblocked. If it is not smooth, it should be cleaned to make it smooth.
If the bearing is damaged or the rotation is blocked, there is a lack of oil in the bearing or there is more dirt in the bearing, which causes the frictional resistance to increase. If the motor does not reach the rated speed, the bearing should be cleaned or replaced in time. There is a layer of oxide on the surface of the commutator. Even if the bump is uneven or the brush does not match the outer edge of the commutator, the rotation speed can be lowered. The commutator and the brush should be cleaned to make the contact good. If the rotor coil is short-circuited or broken, use a multimeter to check and re-wind the coil. When the rotor is in use, the centrifuge tube may be broken, the sample leaks, and the rotor is damaged due to metal fatigue, overspeed, overstress, chemical corrosion, improper selection, unbalanced use of the rotor and temperature loss. For the above situation, the operator is required to be proficient in the operation procedure, correctly select the appropriate centrifuge tube and centrifugal rotor, pay attention to strictly control each step of operation, minimize unnecessary damage caused by human beings, and use it in the safety factor and warranty period of the rotor. .
The power failure is one of the causes of the start of the freezer and the poor cooling effect. The same as the failure of the motor, the power supply and fuse should be checked separately. The voltage is too low, the safety device failure can also make the freezer unable to start, the voltage is too low may be the grid voltage is low; the distribution board wiring is too much; the power line is too long (ideal length is 3 meters). The power line capacity is not enough, which not only reduces the voltage, but also causes accidents. When the power supply voltage drops to 180-190VJ, the freezer cannot start, which affects the cooling effect. The ventilation performance is not good. If the instrument is installed in direct sunlight or poor ventilation, the radiator is not effective, or the radiator is covered with dust, which will affect the cooling effect. The weight of the centrifuge tube is unbalanced and placed asymmetrically; there is foreign matter in the rotor hole to make the load unbalanced; the fixed nut on the upper end of the shaft is loose, the shaft is rubbed or bent; the rotor of the motor does not generate noise in the center of the magnetic field; the damage of the rotor itself can cause the body The vibration is severe and the noise is abnormal.
Most of the above reasons are caused by incorrect operation. After finding out the cause, correct operation can eliminate the abnormal phenomenon. In the course of work, if any abnormality occurs, it should be stopped immediately, and it must not be forced to run, so as to avoid unnecessary losses.
Here
you can find the related products in Transglutaminase For Yogurt, we are professional manufacturer of dairy product, dairy additive. We focused on international export product development,
production and sales. We have improved quality control processes of Transglutaminase For Yogurt to ensure each export qualified product.
If you want to know more about the products in Transglutaminase For Yogurt, please click the product details to
view parameters, models, pictures, prices and other information about Yogurt ingredient, food additive for yogurt.
Whatever you are a group or individual, we will do our best
to provide you with accurate and comprehensive message about Transglutaminase For Yogurt!
Transglutaminase For Yogurt
Transglutaminase for Yogurt ingredient, TG for yogurt, TGase enzyme for yogurt
Guangdong Kelong Biotechnology Co., Ltd. , https://www.kelongfood.com