Human cathepsin K (cath-K) enzyme-linked immunoassay kit instruction manual

Human cathepsin K (cath-K) enzyme-linked immunoassay kit instruction manual


This kit is for research use only .
Detection range: 96T
3pg/ml-150pg/ml

purpose of usage:
This kit is used to determine the content of cathepsin K (cath-K) in human serum, gingival crevicular fluid and related liquid samples.

Experimental principle <br> The kit uses the double antibody sandwich method to determine the level of human proteinase K (cath-K) in the specimen. The microplate was coated with purified human cathepsin K (cath-K) antibody to prepare a solid phase antibody, and cathepsin K (cath-K) was sequentially added to the microcapsules of the coated monoclonal antibody, followed by HRP-labeled tissue. The proteinase K (cath-K) antibody binds to form an antibody-antigen-enzyme-labeled antibody complex, which is thoroughly washed and then added to the substrate TMB for color development. TMB is converted to blue under the catalysis of HRP enzyme and converted to the final yellow color by the action of an acid. The color depth is positively correlated with cathepsin K (cath-K) in the sample. The absorbance (OD value) was measured with a microplate reader at a wavelength of 450 nm, and the concentration of human cathepsin K (cath-K) in the sample was calculated from a standard curve.

Kit composition
1 30 times concentrated washing solution 20ml × 1 bottle 7 Stop solution 6ml × 1 bottle
2 enzyme standard reagent 6ml × 1 bottle 8 standard product (240pg / ml) 0.5ml × 1 bottle
3 Enzyme label coating plate 12 holes × 8 strips 9 standard dilution solution 1.5ml × 1 bottle
4 sample dilution 6ml × 1 bottle 10 instructions 1 copy
5 developer A liquid 6ml × 1 bottle 11 sealing film 2 sheets
6 developer B solution 6ml × 1 bottle 12 sealed bag 1

Specimen requirements
1. The specimens should be extracted as soon as possible after collection, and the extraction should be carried out according to the relevant literature. The experiment should be carried out as soon as possible after extraction. If the test cannot be performed immediately, the specimen can be stored at -20 °C, but repeated freezing and thawing should be avoided.
2. Samples containing NaN3 could not be detected because NaN3 inhibited horseradish peroxidase (HRP) activity.

Human cathepsin K (cath-K) enzyme-linked immunoassay kit instruction manual
1. Dilution of standard: This kit provides one original standard, which can be diluted in a small tube according to the following chart.
120pg/ml Standard No. 5 150μl of the original standard is added to 150μl standard dilution
60pg/ml Standard No. 4 150μl of No. 5 Standard Add 150μl Standard Diluent
30pg/ml Standard No. 3 150μl Standard No. 4 Add 150μl Standard Diluent
15pg/ml Standard No. 2 150μl of Standard No. 3 Add 150μl Standard Diluent
7.5pg/ml Standard No. 1 150μl Standard No. 2 Add 150μl Standard Diluent

2. Adding samples: set blank holes separately (the blank control wells are not added with the sample and the enzyme standard reagent, the other steps are the same), the standard holes, and the sample holes to be tested. Accurately load 50 μl of the standard on the enzyme-labeled plate, add 40 μl of the sample dilution to the well to be tested, and then add 10 μl of the sample to be tested (the final dilution of the sample is 5 times). Add the sample to the bottom of the well of the microplate, try not to touch the wall of the well, and shake gently to mix.
3. Incubation: The plate was sealed with a sealing film and incubated at 37 ° C for 30 minutes.
4. Dosing: 30 times concentrated washing solution diluted with distilled water 30 times and used
5. Washing: Carefully remove the sealing film, discard the liquid, dry it, fill each well with the washing liquid, let stand for 30 seconds, then discard it, repeat 5 times, and pat dry.
6. Add enzyme: 50 μl of enzyme labeling reagent was added to each well, except for blank wells.
7. Incubation: The operation is the same as 3.
8. Washing: The operation is the same as 5.
9. Color development: Add 50 μl of the developer to each well, then add 50 μl of the developer, gently shake and mix, and color at 37 ° C for 15 minutes.
10. Termination: 50 μl of stop solution was added to each well to terminate the reaction (when the blue color turned yellow).
11. Measurement: The absorbance (OD value) of each well was measured sequentially with a blank air conditioner of zero and a wavelength of 450 nm. The measurement should be carried out within 15 minutes after the addition of the stop solution.

Summary of operating procedures:
Calculate the concentration of the standard as the abscissa and the OD as the ordinate. Draw a standard curve on the coordinate paper, and find the corresponding concentration from the standard curve according to the OD value of the sample; multiply by the dilution factor; or use the standard Calculate the linear regression equation of the standard curve by the concentration and OD value, substitute the OD value of the sample into the equation, calculate the sample concentration, and multiply by the dilution factor, which is the actual concentration of the sample.

Precautions
1. The kit should be taken out from the refrigerated environment and should be equilibrated at room temperature for 15-30 minutes before use. If the enzyme label is unsealed after opening the plate, the human cathepsin K (cath-K) enzyme-linked immunosorbent assay kit instruction manual It should be stored in a sealed bag.
2. Concentrated washing liquid may crystallize out. When diluted, it can be heated and dissolved in a water bath. The washing will not affect the result.
3. The sampler should be used for each step, and the accuracy should be corrected frequently to avoid test errors. It is best to control the loading time within 5 minutes. If the number of specimens is large, it is recommended to use a gun.
4. Please make a standard curve at the same time of each measurement, it is best to make a double hole. If the content of the substance to be tested in the specimen is too high (the OD value of the sample is larger than the OD value of the first hole of the standard well), please first dilute the sample dilution with a certain multiple (n times) and then measure it. When calculating, multiply the total dilution by the total dilution. Multiple (×n×5).
5. The sealing film is intended for single use only to avoid cross-contamination.
6. Please keep the substrate away from light.
7. Strictly follow the instructions of the manual, the test results must be based on the microplate reader reading.
8. All samples, washings and various wastes should be treated as infectious materials.
9. The different batch components of this reagent must not be mixed.
10. In the case of an English manual, the English manual shall prevail.

Storage conditions and expiration date
1. The kit is stored at: 2-8 °C.
2. Validity: 6 months

Laser Rangefinder Sensor

Laser rangefinder module, which has two application, one is mainly designed for long distance range finder, another is for secondary development industrial Laser Distance Sensor, which support ttl/usb/rs232/rs485 adapter output data.

With different range measuring program, JRT laser measure tools module, can satisfy customers` different requirements, 200m, 300m, 500m, 1000m, 1200m, 1500m, 3000m.

We have been in this line for 10 years, with a strong R&D ability and hard working, we are now a leading supplier of Laser Distance Meter modules in China.

Product Specifications:

Measuring Range (without Reflection)

5-500m/5-1000m/5-1500m/5-3000m

Accuracy

±1 m

Laser Class

Class I

Laser Type

905nm

Magnification

6X

Battery type

CR2-3V

Operating Temperature

-10-50 ℃ (14-122 ℉ )

Size

28*69*60mm

Weight

About 4g

Operating Temperature

-10-50 ℃ (14-122 ℉ )

Storage Temperature

-25~60 ℃ (-13~140 ℉)


Laser Rangefinder Sensor,Laser Range finder Sensor,Laser Rangefinder Module,Laser Range Finder Module

Chengdu JRT Meter Technology Co., Ltd , https://www.accuracysensor.com